Identification

  • Double stage automatic change-over manifold decompression station for medical gas distribution, max flow rate 40 Nm3/h
  • Double pressure regulation between high pressure source and distribution line
  • Pressure monitoring by means of pressure transducers or pressure switches (optionally)
  • Working manifold sensor integrated in the automatic change-over
  • Provision for integration of alarm module in IP56 cabinet
  • Mounted on painted AISI 430 stainless steel frame
  • Cover panel with window for costant monitoring of outgoing line pressure gauge and alarm module
  • Laser-marked plate indicating production lot and product code

CE marking

Notified body: 0426
CE marking in risk class II B, according to Directive 93/42/EEC

Reference standards

ISO 10524-2: “Pressure regulators for use with medical gases – Part 2: Manifold and line pressure regulators”
ISO 7396-1: ”Medical gas pipeline systems – Part 1: Pipeline systems for compressed medical gases and vacuum”
ISO 15001: “Anaesthetic and respiratory equipment. Compatibility with oxygen”